You must consider some safety precautions to perform fault simulation (also named electrical error simulation) with a SCALEXIO system.
During fault simulation, high currents and voltages might be present on board channels and/or connector pins, which is not expected. This can result in death, personal injury, fire, and/or damage to the SCALEXIO system and connected external devices.
Especially signal measurement channels (such as channels connected in parallel or interconnected reference lines) can carry high currents.
You must ensure that no voltages or currents outside the specified ranges of the I/O channels can occur during fault simulation.
When relays are switched for fault simulation, load rejection can be delayed up to 30 ms due to the switching times of the relays involved. A connected load can be damaged during these 30 ms.
To prevent damage to the relays, switching is normally done when no current load is present. However, a current load cannot be avoided during switching in the following cases:
Switching multiple faults
Load or signal disconnection when switching short circuits between channels
If FRU relays are switched for fault simulation, note the following warnings.
To a varying extent, depending on which loads are connected and which currents and voltages are switched, electric arcs and contact erosion can occur in the FRU relays involved. This will eventually destroy the FRU relays. The board or unit on which they are mounted will no longer be usable and will have to be repaired or replaced.
Operating the relays outside the permitted range (i.e., above the maximum switching capacity) can also destroy the relays and will probably damage the board or unit on which they are mounted.
Before using FRU relays for fault simulation, you must fulfill the following preconditions:
To minimize the risk of damage in a multiple fault scenario, operate the relays only within the permitted conditions and ranges (i.e., under the maximum switching capacity). For concrete values, refer to FRU Relays Data Sheet.
If the connected loads are inductive and you want to simulate faults in their wiring, you must protect the dSPACE hardware from induced high voltages. Refer to Safety Precautions for Using Inductive Loads.
Consider the increased risk of material wear or damage before you set Activation by FRU relay to Allowed in ConfigurationDesk.
Note that defects caused by material wear, misuse or operation outside the permitted ranges are not covered by any warranty, and no liability is accepted by dSPACE for any direct or indirect damage arising from such defects.