There are different methods for powering up and shutting down the LabBox.
The following control elements let you decide how to power up or shut down the LabBox:
Power source switch on the rear side to switch the internal power supply unit of the LabBox on or off.
Automatic Wake-Up switch on the rear side to specify the start behavior of the LabBox.
On/Off button on the front side.
A SCALEXIO AutoBox/LabBox with a DS6001 Processor Board can change its operating mode from standby to normal operating mode when it receives a dedicated wake-up message or frame via a CAN or LIN bus.

To use the wake-up feature, the SCALEXIO system must consist of the following components:
A DS6001 Processor Board installed in a SCALEXIO LabBox.
One or more of the following bus boards:
If multiple LabBoxes are daisy-chained via the power control bus, all the boxes are powered-up when a wake-up message or frame is received by a bus board installed in one of the LabBoxes.
For details, refer to Waking up SCALEXIO AutoBoxes/LabBoxes via Bus Signals.
To set the SCALEXIO LabBox to standby mode via bus signals, you can use the System Shutdown function block. Refer to Basics on Using the System Shutdown Functionality.
The following simplified flow chart displays the different operating modes of the SCALEXIO LabBox and how to switch between them.

Due to internal energy buffers, it can take up to 15 seconds to power down the SCALEXIO LabBox completely after switching off the power source switch. (SCALEXIO LabBox version SY6072-02-xxx requires up to 1 minute to power down.)
Do not establish or remove connections of the power control bus during the powering down period.
Pressing the On/Off button for more than 5 seconds starts a forced shutdown from operating mode to standby mode, which might cause data loss.
The power control bus lets you cascade SCALEXIO AutoBoxes/LabBoxes to power them up or down at the same time by pressing only one of their On/Off buttons. Further, power-up might be performed for all cascaded SCALEXIO AutoBoxes/LabBoxes after a dedicated wake-up message or frame is received at an installed bus board.
To use the power control bus, each SCALEXIO LabBox has a 9-pin, female Sub-D In connector and a 9-pin, male Sub-D Out connector at its rear. You can use the power control bus as shown in the following example to create a daisy chain of SCALEXIO AutoBoxes/LabBoxes.

Observe the following points when using the power control bus:
You can cascade up to 5 SCALEXIO LabBoxes via standard 9-pin Sub-D cables with one-to-one connections from the male (Out) to the female (In) connector.
Do not exceed a cable length of 3 m between the single LabBoxes.
Each LabBox must be directly connected to the power source. (The power control bus does not provide a power connection.)
The LabBox that has an unconnected power control bus In connector works as master. (In the example shown above LabBox 1 is the master.) The setting of the Automatic Wake-Up switch of this master LabBox is valid for all the daisy-chained LabBoxes.
When connecting or disconnecting the power control bus, each LabBox must be switched off by the related power source switch on the rear and shut down properly.
Only line topologies (as shown above) are allowed to connect the power control bus. Do not connect the power control bus in a loop topology.
Do not connect the power control bus to the RS232 connector of a DS6001 Processor Board or vice versa.