Safety Precautions for Working with SCALEXIO LabBox

Potential hazards

WARNING

Risk of death, serious injury, fire, and/or property damage

The SCALEXIO system can carry high currents and high voltages. According to international standards, a voltage higher than 30 VRMS/42.4 VPEAK AC and 60 V DC is classified as hazardous. This presents a risk to people and equipment (death, serious injury, fire, and/or property damage). Therefore, personnel who work with the SCALEXIO system must be informed about the possible dangers and must take suitable safety precautions.


Operating the SCALEXIO system

Note the following points during the operation of the SCALEXIO system:

  • The operator must keep unauthorized people away from the SCALEXIO system by taking suitable safety precautions, for example, locking the system or training the personnel.

  • Use the SCALEXIO system only for measurements of measurement category I. Do not use it for measurements of categories II, III, or IV.

  • The maximum voltage of I/O signals must not exceed 60 V.

  • Operate a SCALEXIO system only with closed enclosures, i.e., all unused slots of the system must be covered by front plates.

  • The operator must not operate a SCALEXIO LabBox in bootloader mode, i.e., using the bootloader mode is only permitted for updating the firmware.

  • Do not work on the external cable harness while the SCALEXIO system is running.

  • Before connecting the SCALEXIO system to the power source, perform a visual inspection of the enclosure and all the connected cables. Do not operate the SCALEXIO system if it looks damaged.

  • Some circuits are live even with the main supply turned off. Before removing or installing I/O boards, switch the power supplies of the host PC and SCALEXIO LabBox off and unplug the power cords. Wait at least one minute to allow all components to discharge. Some of the components, for example, power supply capacitors, can carry residual voltage.

  • The On/Off button at the front of SCALEXIO LabBox does not disconnect the SCALEXIO system from the mains. For complete disconnection, unplug the power cord from the socket.

  • The SCALEXIO hardware provides electrical energy at the I/O pins, which can cause a fire if external components such as sensors/actuators are not appropriately connected. To prevent a fire, apply the general fire safety regulations, e.g., supervise the operation, remove fire loads, and use fire-proof materials and enclosures.

  • High voltages can be present at pins of the ECU/load connectors. Therefore, do not leave the connectors unconnected during operation.

  • High currents can be present at connector pins or interfaces where they are not expected, for example, due to incorrect external wiring.

  • If you measure analog signals with unshielded I/O cables, do not operate mobile phones, hand-held transceivers, or any other sources of electromagnetic fields close to the cable harness of the SCALEXIO system during run time. Otherwise, you might influence the measurements.


Using measurement instruments

To avoid an electric shock and/or property damage, note the following points when you work with measurement instruments (for example, oscilloscope or tester device):

  • Use only suitable and tested measuring instruments. Observe all the safety instructions when making measurements.

  • Connect and disconnect the measuring instruments only when the SCALEXIO system is turned off.

  • Before starting measurements, check that the measuring instruments and their cables are in perfect condition.


Unexpected behavior of the SCALEXIO system

If the miniature circuit breaker or residual current device switches off the SCALEXIO system, disconnect the system from the power source immediately and make sure that it cannot be switched on unintentionally. The SCALEXIO system must not be put into operation again until it has been analyzed, repaired, and approved by dSPACE or an authorized support engineer.


Protecting the SCALEXIO system

  • Before a lightning storm, disconnect the LAN and power cords. Alternatively, install appropriate protection devices.

  • Do not unplug the connectors by pulling the cables. Hold the plug itself to pull it out.

  • Route external cables so that they are not likely to be walked on or pinched by items placed upon or against them.


Cleaning the SCALEXIO system

  • Before cleaning, disconnect the SCALEXIO system from the power source.

  • Do not use aerosol cleaners, water, or other liquids to clean the HIL system. Instead, use a damp cloth for cleaning.