You must take the following safety precautions to avoid risks of injury when you work with the MicroLabBox II.
Using the MicroLabBox II in a humid environment might result in electric shock due to hazardous voltages or might damage the MicroLabBox II.
According to IEC 61010-1 (product safety), use the MicroLabBox II only in a dry environment and avoid condensation.
Do not allow the MicroLabBox II to come into contact with liquids.
Movable devices and applied voltages can cause injuries and/or hardware damage.
Use designated and protected test areas with an emergency shutdown mechanism if moving devices could cause injury or hardware damage.
Keep in mind that unexpected movements of moving devices are possible if no application is running or if an incorrect model is used, for example.
Make sure that connected actors remain in a defined position if no real-time application is running on the the MicroLabBox II, for example, by connecting external pull-up and pull-down resistors.
Output channels are set to high impedance until the real-time application starts and after the real-time application is unloaded. From the time the real-time application stops until the real-time application is unloaded, the stop values are active.
Before you connect an external device to the MicroLabBox II, make sure that the applied voltages do not exceed the protected voltage ranges of the MicroLabBox II. Keep in mind that connected devices might provide unexpected hazardous voltages.
Set up the the MicroLabBox II in such a way that you can quickly and safely switch off the mains, e.g., by quickly accessing the mains switch.
If you intend to operate the the MicroLabBox II without continuous supervision, observe the following safety instructions:
The building in which the the MicroLabBox II is operated must have a working fire alarm system.
Unattended operation is only permitted in automatic mode.
The devices under test (DUT) must be suitable for continuous operation.
Comply with any legal provisions and regulations for building safety applicable in the country of use.
Opening the housing and modifying the MicroLabBox II might result in electric shock due to hazardous voltages.
Open the bottom flap only if the MicroLabBox II is disconnected from the mains.
If you replace the power cord, make sure that the replacement power cord fulfills the mains input characteristics of the MicroLabBox II.